5. General Safety Instructions

Medi Client IIA - Instructions for use (Version 1.03)

Note
The Medi Client IIA is intended for use in the electromagnetic environments specified below. The user of this equipment
should assure that is used in such an environment.

Electromagnetic Emissions

Medi Client IIA is intended for use in the electromagnetic environment specified below. The operator should assure that it is used in
such an environment.

Emissions Compliance

according

to... Electromagnetic

environment

RF emissions (CISPR 11)

Group1

The Medi Client IIA uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.

CISPR Emissions Classification

Class B

Harmonic emissions
(IEC 61000-3-2)

Class D

Voltage fluctuations / flicker
(IEC 61000-3-3)

Complies

The Medi Client IIA is suitable for use in establishments
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.

Table 1: Information regarding electromagnetic emissions (IEC 60101-1-2, table 201)

Electromagnetic Immunity

The Medi Client IIA is intended for use in the electromagnetic environment specified below. The operator should assure that it is used
in such an environment.

Immunity against...

IEC 60601-1-2 test level

Compliance level
(Medi Client IIA)

Electromagnetic environment

electrostatic discharge,
ESD

(IEC 61000-4-2)

contact discharge: ± 6 kV

air discharge: ± 8 kV

± 2, 4, 8 kV,

except for the interfaces
marked with the ESD symbol

Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.

electrical fast
transients / bursts

(IEC 61000-4-4)

power supply lines: ± 2 kV

longer input / output lines:

± 1 kV

± 2kV,

± 1 kV

Mains power quality should be that of a
typical commercial or hospital environment.

surges on AC mains
lines

(IEC 61000-4-5)

Common mode: ± 2 kV

differential mode: ± 1 kV

± 2kV,

± 1 kV

Mains power quality should be that of a
typical commercial or hospital environment.

power frequency
magnetic field 50/60
Hz
(IEC 61000-4-8)

3 A/m

30 A/m

If high levels of power frequency magnetic
fields occur, the system maybe influenced, so
therefore avoid system operating nearby
AC-mains cable runway or cable bundle.

voltage dips and short
interruptions on AC
mains input lines

(IEC 61000-4-11)

>95% Interruption for 10 ms
30% Dips for 500ms
60% Dips for 100ms
>95% Interruption for 5000ms

according the test levels from
IEC60601-1-2

Mains power quality should be that of a typical
commercial and/or hospital environment. If the
user of the Medi Client IIA requires continued
operation during power mains interruptions, it is
recommended that the Medi Client IIA is powered
from an uninterruptable power supply.

Table 2: Electromagnetic Immunity