Bio-Rad Microplate Manager Software User Manual

Page 41

background image

Chapter 4. Templates

31

Customized name for each assay.

Notes about each assay to be printed on reports.

Type of regression for each assay.

Type of axes for each regression.

External standards (if any) to be applied to each assay

Types of Wells

A template allows you to define six different types of wells:

Standard wells contain your standards. They are indicated by a circle with a
number at the center that indicates the replicate group.

Unknown wells contain your unknowns. They are indicated by a square with a
number at the center that indicates the replicate group.

Blank wells are used to measure background noise for accurate absorbance
calculation. They are indicated by a diamond with a letter B at the center.

Positive control wells contain samples that define the upper limit of your data
range. They are indicated by a triangle with a “+” at the center.

Negative control wells contain samples that define the lower limit of your data
range. They are indicated by an inverted triangle with a “–” at the center.

Undefined wells are undefined in the template. They are indicated by an empty
white square.