beautypg.com

Bio-Rad Experion RNA Analysis Kits User Manual

Page 53

background image

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

49

Fig . A .1 . Electropherogram generated by an Experion RNA HighSens analysis of mRNA . The 18S rRNA peak is identified by
Experion software.

The appearance of the electropherogram depends on the type of RNA assay performed:

n

mRNA analyses often yield broad peaks or a smear and, on occasion, contaminating 18S

and 28S rRNA peaks (Figure A.1). In addition, mRNA samples require more time for analysis
than total RNA samples to ensure inclusion of the population of longer transcripts and to
achieve a smooth baseline

n

Total RNA analyses yield 2–3 prominent peaks corresponding to rRNA (18S and 28S and

occasionally 5S in eukaryotic sources, 16S and 23S in prokaryotic sources, Figure A.2)

5S

Fig . A .2 . Electropherogram generated by an Experion RNA HighSens analysis of eukaryotic total RNA . The 18S and 28S
rRNA peaks are identified by Experion software.

Experion RNA StdSens and HighSens Analysis Kits