beautypg.com

3 contacting technical support – Bio-Rad Experion RNA Analysis Kits User Manual

Page 49

background image

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

45

Error

Probable Cause

Recommended Action

Lower marker is identified
incorrectly

Experion software selected incorrect peak to
correspond to the marker; such an error affects
alignment and sizing

Manually select the markers by following the instructions in
Section 6.1

Virtual gel does not
resemble an agarose gel
separation of same samples

Gel view is a direct interpretation of
electropherogram data. By default,
fluorescence intensity of bands in each lane is
scaled to intensity/peak height of highest peak
in that sample (scale to local). Consequently,
each gel lane may have a different scale of
fluorescence intensity

Adjust scaling of the virtual gel to Scale To Global (see
Section 4.2). This adjusts the scale on all electropherog
rams

Adjust the intensity of the virtual gel using the sliding cursor

Baseline is fluctuating (for
example, into very negative
and positive values)

Air bubbles are interfering with electrical
contact in one or more of the wells

Stop the run, remove the chip, and remove and replace
solution in affected wells

Sample is in inappropriate sample buffer

Ensure RNA is in buffer of low ionic strength (for Experion
RNA StdSens use TE or water, for Experion RNA HighSens
use water)

Chip is not completely primed

Prime and load a new chip

Samples are too concentrated; RNA levels are
out of range of assay

Dilute the samples and repeat the analysis

RNA ladder or samples
appear degraded

RNase contamination of electrodes, reagents,
or consumables

Improper storage

Perform the deep cleaning procedure described in
Appendix B and software Help section (search term
“electrodes”)

Use only 0.2 µm-filtered water (such as Experion DEPC-
treated water) for diluting samples and the RNA ladder

Use RNase-free consumables and RNase-free, filter pipet
tips; change pipet tips with each use

Store RNA ladder and samples appropriately (for example,
at –70°C)

7.3 Contacting Technical Support

If you are having problems with a run and would like the Bio-Rad Technical Support team to review your
data, please submit the following files to LSG_TechServ_US@bio-rad .com for analysis:

n

Run file (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\

name>\.bdf)

n

Run log (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\

name>\.log)

n

Packet file (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\

name>\Packet\.txt)

n

System log (C:\Program Files\Bio-Rad Laboratories\Experion Software\Backup\Version X\

Experion_System.log or .txt)

n

Lot numbers of the RNA chips and the Experion RNA analysis kit

To export the run file, log file, and packet file, select the run(s) in the tree view and click Export Selected
Runs. The software exports the files into folders with the corresponding project name(s).

Experion RNA StdSens and HighSens Analysis Kits