beautypg.com

Experion rna stdsens and highsens analysis kits – Bio-Rad Experion RNA Analysis Kits User Manual

Page 47

background image

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

43

Error

Probable Cause

Recommended Action

No peaks detected in
some samples

There is not enough sample in the wells

Ensure pipets are calibrated and that correct volumes have
been added to the wells

Samples were not prepared properly

Review the sample preparation procedure detailed in
Chapter 3

RNA stain was photobleached

Protect RNA stain and gel-stain solution (GS) from light

Prepare new GS; use a new tube of RNA stain if necessary

Particulates are clogging the microchannels

Use only high-quality 0.2 µm-filtered water (such as
Experion DEPC-treated water, not autoclaved water)

Verify that gel (G) and GS were properly filtered

If using samples that contain particulates, perform a quick
microcentrifuge spin of the prepared sample to pellet
particulates before loading

Remove solutions from the top of the liquid to prevent
introducing particulates pelleted during centrifugation

No peaks detected in any
samples

Gel-stain solution (GS) is old or photobleached

Prepare fresh GS

Chip was primed with incorrect pressure and
time settings

Check settings on the priming station and repeat analysis
with a new chip

Particulates are clogging microchannels

Use only high-quality, 0.2 µm-filtered water (such as
Experion DEPC-treated water, not autoclaved water)

Verify that gel (G) and GS were filtered properly

If using samples that contain particulates, perform a quick
microcentrifuge spin of the prepared sample to pellet
particulates before loading

Remove solutions from top of liquid to prevent introducing
particulates pelleted during centrifugation

Lower marker is missing

Samples do not contain loading buffer or were
not prepared properly

Review the sample preparation procedure in Chapter 3

Electropherogram peaks are
much smaller than expected
(peaks are <50% of what is
expected)

Not enough sample was added

Ensure pipets are calibrated to accurately deliver ≤10 µl

Too much salt in sample; salt ions compete
with sample ions during injection

Ensure RNA is in buffer of low ionic strength (for RNA
StdSens use TE or water; for RNA HighSens use water)

Analyze a dilution series of sample in water on another
chip. If peak heights and concentrations of diluted sample
increase, salt is too high in the original sample. Determine
proper dilution to minimize salt effect

Gel-stain solution (GS) is old or photobleached

Prepare fresh GS

Reagents, specifically RNA stain, were not
brought to room temperature prior to use; if
stain is not equilibrated to room temperature, it
will be more concentrated when it is added to
gel, which may result in lower peaks and areas

Repeat analysis with a new chip, new tube of stain, and
properly equilibrated reagents

Filtered gel (G) or gel-stain solution (GS) were
mislabeled or used improperly. RNA chips
contain wells for GS that carry sample through
microchannels and supply dye for analysis

Ensure chip was primed with GS

Ensure G was applied to well labeled G and GS was used in
all wells marked GS

Experion RNA StdSens and HighSens Analysis Kits