Bio-Rad GS-900™ Calibrated Densitometer User Manual
Page 62
Vertical Streaking
(contd.)
Problem Cause
Solution
Twin vertical spots
Improper placement of the IPG
Make sure that the focused IPG strip is
or vertical doublets
strip onto the gel
in full contact with the gel
Temperature gradient in the gel
Lower the power settings for the second-
dimension SDS-PAGE run, especially when
using cells that provide only one-sided
cooling of the gel
Use a better circulation system to improve
heat dissipation during a run
Blank vertical stripes
Air bubble trapped in the agarose Ensure that the 2-D gel has a straight,
that joins the IPG strip to the top
level top edge and that the IPG strip is
of the gel
in direct contact with the 2-D gel along its
entire length. Squeeze out air bubbles
by pressing on the plastic backing of
the IPG strip
Use a 0.5% agarose overlay solution to
prevent the IPG strip from coming loose
or moving. To minimize the number of
bubbles in the overlay, melt the agarose
overlay solution completely prior to loading
Insufficient rehydration of a region Make sure that the IPG strip is not sticking
of the IPG strip, or tears resulting to the bottom of the rehydration tray
from improper handling, resulting
in the absence of focused protein Check the integrity of rehydrated IPG
in that region
strips prior IEF
Focusing of an amphoteric
Apply sample cleanup
nonprotein contaminant
(for example, phospholipid or
HEPES) prevents protein focusing
around the pI of the contaminant
Vertical Streaking
(contd.)
Problem Cause
Solution
Blank stripes near pH 7
Excessive DTT (>50 mM) in the
Lower the amount of DTT in the
IPG sample solution
rehydration solution
Blank stripes at the
Salt buildup
Remove ionic contaminants from the
electrodes, especially
samples with Bio-Rad´s ReadyPrep 2-D
at the cathode
cleanup kit or by desalting
Blank vertical regions
Interfering substances; impurities Remove contaminants from the samples
in the rehydration/sample solution with the ReadyPrep 2-D cleanup kit or
by desalting
Use high-quality reagents and chemicals
for electrophoresis to minimize the risk
of impurities. Replace chemicals of
questionable or unknown shelf life, origin,
or quality, as these products can also
contribute to poor 2-D results
Air bubble trapped in the agarose Ensure that the 2-D gel has a straight,
that joins the IPG strip to the top
level top edge and that the IPG strip is
of the gel
in direct contact with the 2-D gel along
its entire length. Squeeze out air bubbles
by pressing on the plastic backing of
the IPG strip
Use a 0.5% agarose overlay solution to
prevent the IPG strip from coming loose or
moving. To minimize the number of bubbles
in the overlay, melt the agarose overlay
solution completely prior to loading
120
121
2-D Electrophoresis Guide
Troubleshooting