beautypg.com

Bio-Rad GS-900™ Calibrated Densitometer User Manual

Page 61

background image

Horizontal Streaking

(contd.)

Problem Cause

Solution

Spots

Incomplete IEF

Optimize the sample focusing time by
running a time course. For example,
run the sample on 6 IPG strips and
remove an IPG strip at each time point
(20 kV-hr, 30 kV-hr, 40 kV-hr, etc.)


Intermittent

Contaminants such as salts,

Use appropriate contaminant removal

ionic detergents (for example,

techniques, such as treatment with

SDS), peptides, nucleic acids,

the ReadyPrep 2-D cleanup kit

lipids, polysaccharides,

phenolic compounds








Vertical Streaking

Across the entire gel

Leaking of the upper buffer

Prior to inserting the gel(s) into the vertical

reservoir (cathode) of the vertical

electrophoresis cell, wet the gaskets of the

electrophoresis unit

electrophoresis chamber with water or
use a small amount of vacuum grease




Incomplete equilibration

Increase equilibration time to 15 min

Old DTT and iodoacetamide

Use fresh reagents for the equilibration step

preparations used in equilibration

Vertical Streaking

(contd.)

Problem Cause

Solution

At one end of the gel

Protein aggregation or

Dilute the sample to 3–5 µg/µl for

(cup loading)

precipitation caused by

cup loading

too much protein or sample

loading problems

Perform a protein assay prior to IEF to
ensure correct protein load. The total
amount of protein that should be loaded
onto an IPG strip depends on the length
of the strip and the stain that will be used
to visualize the results

Load the sample using in-gel sample loading

Prolong the time on the initial low-voltage
steps and increase the voltage gradually

Field strength used for sample

Reduce the field strength to ~10 V/cm

loading is too high

IPG strip length

Poor protein solubility

Increase the solubilizing strength of
2-D sample solution

Isolated streaking

Improper rehydration of IPG strip

Check the rehydration volumes and times
for the lengths of IPG strips used

If the sample appears unevenly distributed,
or if areas of the IPG strip are not wetted
with sample, slide the IPG strip back and
forth several times along the length of the
channel in the focusing tray


Point streaking

Dust or other particles in the

Filter gel solutions through a 0.45 μm

(handcast gels)

gel solutions

membrane and into a dust-free container

Vertical streaks

Insufficient binding of SDS

Check the SDS concentration (>1%) in the

connected to a spot

to protein

equilibration solution

Increase equilibration time:

equilibrate IPG strips for 2 × 15 min

Incorrect pH in resolving gel buffer; Ensure that the pH of the Tris buffer used

incorrect pH decreases mobility

for gel casting is 8.8

of protein-SDS complexes and

causes vertical streaks

Buffer leakage

Ensure that the upper buffer reservoir
is not leaking

118

119

2-D Electrophoresis Guide

Troubleshooting