Bio-Rad GS-900™ Calibrated Densitometer User Manual
Page 61
Horizontal Streaking
(contd.)
Problem Cause
Solution
Spots
Incomplete IEF
Optimize the sample focusing time by
running a time course. For example,
run the sample on 6 IPG strips and
remove an IPG strip at each time point
(20 kV-hr, 30 kV-hr, 40 kV-hr, etc.)
Intermittent
Contaminants such as salts,
Use appropriate contaminant removal
ionic detergents (for example,
techniques, such as treatment with
SDS), peptides, nucleic acids,
the ReadyPrep 2-D cleanup kit
lipids, polysaccharides,
phenolic compounds
Vertical Streaking
Across the entire gel
Leaking of the upper buffer
Prior to inserting the gel(s) into the vertical
reservoir (cathode) of the vertical
electrophoresis cell, wet the gaskets of the
electrophoresis unit
electrophoresis chamber with water or
use a small amount of vacuum grease
Incomplete equilibration
Increase equilibration time to 15 min
Old DTT and iodoacetamide
Use fresh reagents for the equilibration step
preparations used in equilibration
Vertical Streaking
(contd.)
Problem Cause
Solution
At one end of the gel
Protein aggregation or
Dilute the sample to 3–5 µg/µl for
(cup loading)
precipitation caused by
cup loading
too much protein or sample
loading problems
Perform a protein assay prior to IEF to
ensure correct protein load. The total
amount of protein that should be loaded
onto an IPG strip depends on the length
of the strip and the stain that will be used
to visualize the results
Load the sample using in-gel sample loading
Prolong the time on the initial low-voltage
steps and increase the voltage gradually
Field strength used for sample
Reduce the field strength to ~10 V/cm
loading is too high
IPG strip length
Poor protein solubility
Increase the solubilizing strength of
2-D sample solution
Isolated streaking
Improper rehydration of IPG strip
Check the rehydration volumes and times
for the lengths of IPG strips used
If the sample appears unevenly distributed,
or if areas of the IPG strip are not wetted
with sample, slide the IPG strip back and
forth several times along the length of the
channel in the focusing tray
Point streaking
Dust or other particles in the
Filter gel solutions through a 0.45 μm
(handcast gels)
gel solutions
membrane and into a dust-free container
Vertical streaks
Insufficient binding of SDS
Check the SDS concentration (>1%) in the
connected to a spot
to protein
equilibration solution
Increase equilibration time:
equilibrate IPG strips for 2 × 15 min
Incorrect pH in resolving gel buffer; Ensure that the pH of the Tris buffer used
incorrect pH decreases mobility
for gel casting is 8.8
of protein-SDS complexes and
causes vertical streaks
Buffer leakage
Ensure that the upper buffer reservoir
is not leaking
118
119
2-D Electrophoresis Guide
Troubleshooting