Isoelectric focusing – Bio-Rad GS-900™ Calibrated Densitometer User Manual
Page 58
SDS-PAGE
Problem Cause
Solution
Low or zero current,
With a precast gel, the tape
Remove the tape
and samples do not
at the bottom of the gel cassette
migrate into the gel
was not removed
Insufficient buffer in the inner
Fill the inner and outer buffer chambers
or outer buffer chamber
to ensure that the IPG well is
completely covered
Electrical disconnection
Check the electrodes and connections
Running time slower or
Incorrect running buffer
Check the buffer composition and type
faster than expected
concentration or type
Leaking from inner
Incomplete gasket seal
Wet the gasket with running buffer
buffer chamber
before use
Improper assembly of the gel
■
Ensure that the top edge of the short
into the electrode/ companion
plate fits under the notch at the top
assembly
of the gasket
■
Ensure that the top of the short plate
touches the green gasket
Isoelectric Focusing
Problem Cause
Solution
Initial low or zero current
Poor contact between IPG strips
Make sure that the gel side of the IPG strip
and electrodes
is in contact with the electrode
For the gel-side down configuration with
the PROTEAN
®
i12
™
cell, use the IPG
strip retainers
Incomplete wetting of
Wet the electrode wicks with distilled or
electrode wicks
deionized H
2
O until they are damp, but not
soaking wet
Incomplete IPG strip rehydration
Check the rehydration volumes and times
for the lengths of IPG strips used
No current in any lane
No contact between the
Make sure that:
electrode assembly and
■
The electrode assembly is properly
IPG strips
seated in the focusing tray
■
The IPG strips are positioned correctly,
(for example, that the gel is in direct
contact with the electrode)
No contact between the
Make sure that:
electrode assembly
■
The gold contact pin of the negative (–)
and instrument
assembly is in direct contact with the
cathode bar on the instrument
■
The positive (+) assembly is completely
inserted into the anode of the instrument
Voltage does not
High levels of ionic contaminants
Keep salt concentrations under 40 mM;
increase beyond initial
in sample solution (optimum salt
if necessary, desalt the sample
low voltage steps
concentration is ~10 mM, though
(for example with Micro Bio-Spin
™
6
up to 40 mM can be tolerated)
columns or the ReadyPrep
™
2-D cleanup kit)
Salt collects in electrode wicks, so
replace electrode wicks from time to time
(every 2 hr) during the initial low-voltage
steps. Several hours may be needed for
ionic contaminants to leave IPG strips
Voltage does not reach
Programmed voltage is too
Lower the voltage maximum set for the
programmed value,
high for the pH range and
focusing step; the conductivity and the
or maximum voltage is
length of IPG strip
length and type of IPG strip determine the
reached very slowly.
voltage maximum that can be reached
Note: good focusing
may be obtained even
if programmed voltage
is never reached
Ampholyte concentration is
Lower the ampholyte concentration
too high. Up to 1% (v/v)
Bio-Lyte
®
ampholytes may
be used, but ampholytes
increase conductivity;
therefore, voltage will be lower
with increasing concentrations
Isoelectric Focusing
(contd.)
Problem Cause
Solution
Excess sample during rehydration Use correct rehydration volumes for
did not enter gel, or IPG gels are
the lengths of the IPG strips used
overswelled with excess sample
Voltage is too high for the IPG
Program Vh for the IEF step to ensure
strip size and pH gradient
complete focusing of the sample
Large fluctuations in
IPG strips contain poorly
Check rehydration volumes and times
voltage and current
rehydrated regions, or IPG strips
have dried out during the run
Make sure that the rehydration solution is
evenly distributed during rehydration and
that the IPG strips are completely covered
with mineral oil
Burning of strips
Current limit is too high
Use a current limit of 50 µA/IPG strip
IPG strips have dried out
Make sure that the IPG strips are covered
with mineral oil or equivalent
Electrode wicks are too wet or
Wet the electrode wicks with distilled
contain incorrect electrode solution or deionized H
2
O until they are damp,
not soaking wet
Incorrect rehydration
Check the composition of the
solution composition
rehydration solution
Sample is leaking from
Cup positioning is incorrect
When positioning the cup holder, make
the sample cups
sure that it clicks into place at the edges
of the focusing tray
The cup is positioned in an
Make sure that the IPG strips are
area of the IPG strip that is
rehydrated evenly and thoroughly
not completely rehydrated
The cup is malfunctioning
Replace cup
112
113
2-D Electrophoresis Guide
Troubleshooting