Total protein staining, D gel evaluation – Bio-Rad GS-900™ Calibrated Densitometer User Manual
Page 59
Total Protein Staining
Problem Cause
Solution
Spots not visible (see 2-D
No protein in the gel
Use another staining method to confirm
Gel Evaluation, below)
that there is protein in the gel
Malfunctioning imaging
Check the instrument manual for
system or incorrect
troubleshooting information, or contact
imaging parameters
the imaging instrument manufacturer
Poor staining
Insufficient protein in the gel
Repeat the experiment with a
load sensitivity
higher protein quantity
Dirty staining trays (for example,
Clean the staining trays and other
with silver staining)
equipment thoroughly with laboratory
glassware cleaner
Insufficient stain volume
Follow the recommendations for stain
volume appropriate to the gel size
Insufficient staining time
Increase staining time
Reuse of staining solution
To ensure quantitative reproducibility
of a 2-D experiment, never reuse
staining solution
High or uneven
Dirty equipment or staining trays
Clean the staining trays and other equipment
background staining
thoroughly with laboratory glassware cleaner
Too much time in staining solution Restrict the time in staining solution
as recommended
Wash the gel in water or respective
destaining solution for >30 min
Reagent impurities
Make sure that the water and reagents
used for staining are of the highest
possible quality
Diffuse, uneven
Insufficient washing
Perform more washing steps. Use purified
background in
laboratory water and clean staining trays
silver-stained gel
Do not place too many gels in one tray.
Fully immerse the gels in the staining
solution; they should not stick to the
staining tray
Insufficient fixative (some uneven
Apply a longer fixing procedure
background stain is normal when
using a silver stain. Due to migration
of different chemicals and ions
into the gels, some regions can
be stained with different colors
or intensities)
Contaminant(s) in the agarose
Prepare fresh overlay solution
overlay solution
Total Protein Staining
(contd.)
Problem Cause
Solution
Speckles or blotches in
Particulate material from reagents, Clean the staining trays and other
the gel image
staining tray, dust, or gloves
equipment thoroughly with laboratory
glassware cleaner
Limit exposure of gels and staining
solution to open air
Use dust-free gloves, and handle gels
only by the edges
Uneven staining
Insufficient shaking during staining Agitate the gel during staining
Gel shrinkage
Some gel shrinkage occurs
Transfer the gel to water
during staining
2-D Gel Evaluation*
Problem Cause
Solution
No Spots or Fewer Spots
than Expected
Across the gel
Insufficient sample was loaded
Check the sample concentration by
protein assay
Check that the protein assay is functioning
properly and that it is not responding to
interfering substances in your sample
Insufficient sample entered
Start IEF at a low field strength
the IPG strip
Make sure that the IPG strips are in the
correct orientation in the focusing tray
Check that the orientation of
electrical connections
Increase the solubility strength of the
2-D sample solution; insoluble proteins
will not enter the IPG strip
Failure of detection reagents
Run a lane of unstained standards adjacent
to the second-dimension separation. If the
standards are not detected, check the
expiration dates and the formulations of
all detection reagents
Staining method not
See Chapter 3 for sample loading
sensitive enough
recommendations dependent on the
staining technique used
Poor protein transfer from
Perform the first stage of SDS-PAGE at
IPG strip to SDS-gel
low voltage (50 V) for >20 min until the
bromophenol blue front enters the
separation gel (time depends on gel size)
* Also refer to Berkelman et al. (2004) and Bio-Rad Laboratories (2005).
114
115
2-D Electrophoresis Guide
Troubleshooting