beautypg.com

Total protein staining, D gel evaluation – Bio-Rad GS-900™ Calibrated Densitometer User Manual

Page 59

background image

Total Protein Staining

Problem Cause

Solution

Spots not visible (see 2-D

No protein in the gel

Use another staining method to confirm

Gel Evaluation, below)

that there is protein in the gel

Malfunctioning imaging

Check the instrument manual for

system or incorrect

troubleshooting information, or contact

imaging parameters

the imaging instrument manufacturer

Poor staining

Insufficient protein in the gel

Repeat the experiment with a

load sensitivity

higher protein quantity

Dirty staining trays (for example,

Clean the staining trays and other

with silver staining)

equipment thoroughly with laboratory

glassware cleaner

Insufficient stain volume

Follow the recommendations for stain

volume appropriate to the gel size

Insufficient staining time

Increase staining time

Reuse of staining solution

To ensure quantitative reproducibility

of a 2-D experiment, never reuse

staining solution

High or uneven

Dirty equipment or staining trays

Clean the staining trays and other equipment

background staining

thoroughly with laboratory glassware cleaner

Too much time in staining solution Restrict the time in staining solution

as recommended

Wash the gel in water or respective

destaining solution for >30 min

Reagent impurities

Make sure that the water and reagents

used for staining are of the highest

possible quality

Diffuse, uneven

Insufficient washing

Perform more washing steps. Use purified

background in

laboratory water and clean staining trays

silver-stained gel

Do not place too many gels in one tray.

Fully immerse the gels in the staining

solution; they should not stick to the

staining tray

Insufficient fixative (some uneven

Apply a longer fixing procedure

background stain is normal when

using a silver stain. Due to migration

of different chemicals and ions

into the gels, some regions can

be stained with different colors

or intensities)

Contaminant(s) in the agarose

Prepare fresh overlay solution

overlay solution

Total Protein Staining

(contd.)

Problem Cause

Solution

Speckles or blotches in

Particulate material from reagents, Clean the staining trays and other

the gel image

staining tray, dust, or gloves

equipment thoroughly with laboratory

glassware cleaner


Limit exposure of gels and staining

solution to open air


Use dust-free gloves, and handle gels

only by the edges

Uneven staining

Insufficient shaking during staining Agitate the gel during staining

Gel shrinkage

Some gel shrinkage occurs

Transfer the gel to water

during staining

2-D Gel Evaluation*

Problem Cause

Solution

No Spots or Fewer Spots
than Expected

Across the gel

Insufficient sample was loaded

Check the sample concentration by

protein assay

Check that the protein assay is functioning

properly and that it is not responding to

interfering substances in your sample

Insufficient sample entered

Start IEF at a low field strength

the IPG strip

Make sure that the IPG strips are in the

correct orientation in the focusing tray

Check that the orientation of

electrical connections

Increase the solubility strength of the
2-D sample solution; insoluble proteins
will not enter the IPG strip

Failure of detection reagents

Run a lane of unstained standards adjacent
to the second-dimension separation. If the
standards are not detected, check the
expiration dates and the formulations of
all detection reagents

Staining method not

See Chapter 3 for sample loading

sensitive enough

recommendations dependent on the

staining technique used

Poor protein transfer from

Perform the first stage of SDS-PAGE at

IPG strip to SDS-gel

low voltage (50 V) for >20 min until the

bromophenol blue front enters the

separation gel (time depends on gel size)

* Also refer to Berkelman et al. (2004) and Bio-Rad Laboratories (2005).

114

115

2-D Electrophoresis Guide

Troubleshooting