beautypg.com

ADS Environmental Services XiLog+ QR 775026 A0 User Manual

Page 70

background image

ADS XiLog+ System User Manual

64

 Select the interval at which the data will be logged and averaged (as

necessary) from the Interval drop-down list. Sending 5- or 10-
minute data will require sending more than one SMS per channel
per day. Choosing Off will prevent the channel from recording
measurements.

5. Designate the parameters under the Pressure Channel 1 header in the

following way:

 Select the unit of measure in which to log the data from the Units

drop-down list.

 Select the frequency at which to record the data on this channel

from the Interval drop-down list. Choose Off to prevent the
channel from recording data.

6. Initiate logger programming based on the designated parameters in one

of the following ways:

Program Logger, Delete Data To delete all old data from the

logger during programming, click on the Start button.

Program Logger, Retain Data To program the logger without

deleting the data, click on the down arrow to the right of the Start
button and select the Changes Only option.

XiLog+ FastTrack programs the logger based on the selected method.
Pressing start also synchronizes the logger clock to the PC clock.