beautypg.com

5 appendix, 1 technical data – Metrohm 793 IC Sample Preparation Module User Manual

Page 42

background image

5 Appendix

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

38

5 Appendix

5.1 Technical data

Pump

Pump type

2-channel peristaltic pump with rotational speed

of

20/min (50 Hz)

24/min (60 Hz)

Pump capacity

typ. ~1.3 mL/min with 6.1826.010 pump tubing

Pressure

max. 4 bar (0.4 MPa)

Pumpable liquids

Clear liquids with no solid contents

Pump tubing material

PVC (Tygon

£

)

Remote interface for pump

Purpose

Remote control of pump (switching on/off)

Pin assignment

330

+5V

2

5

+

-

Pin

Remote interface for SP module A

Purpose

Remote control of actuator (rotating)

Pin assignment

Pin

1k

+5V

2

5

+

-

10k

0V

* A precondition for triggering a step is that the

current flows for at least 10 ms. In order that a

new step can be triggered no current must flow

for at least 10 ms before.

Current flows:

on

Current does not flow: off

external

793

Current flows:

Step

*

Current does not flow: –

external

793