2.1 Setting up the instrument

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

7

2 Installation

2.1 Setting up the instrument

2.1.1 Packaging

The 793 IC Sample Prep Module is supplied together with the sepa-

rately packed accessories in special packagings containing shock-

absorbing foam linings designed to provide excellent protection. The

actual instrument is packed in an evacuated polyethylene bag to pre-

vent the ingress of dust. Please store all these special packagings as

only they can assure damage-free transport of the instrument.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and

has arrived without damage (compare with delivery note and list of

accessories in section 5.2). In the case of transport damage, see

instructions in section 5.4.1 "Warranty".

2.1.3 Location

Position the instruments in the laboratory at a location convenient for

operation, free from vibrations and protected against a corrosive

atmosphere and contamination by chemicals.

2.1.4 Arrangement of the instruments

The best position for the 793 IC Sample Prep Module is beside the Mo-

dular IC System (e.g. 733 IC Separation Center).

The 793 IC Sample Prep Module should always be placed at the

bottom of the instrument stack so that any leaks which may occur in

the pump tubing or connections cannot cause damage to the other

instruments by leakage of corrosive liquids (e.g.acid).