2.3 Connection to a Modular IC System for Neutralization

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

11

Please note, that the suppressor of the 753 IC Suppressor Module is

not switched by the 766 Sample Processor program. The remote con-

nection 16

16

16

16 (suppressor/actuator) of the 753 IC Suppressor Module is

connected to the event line “fill” of the 733 valve B instead, using the

6.2128.180 cable. Thus the suppressor is switched when the fill position

is accessed by valve B.

2.3.2

Connection of the SP Module A

The following description implies the usage of the 2.733.0120 IC Sepa-

ration Center (two injectors).

The 1.793.0110 SP Module A must first be inserted in the 733 IC Sepa-

ration Center and connected to the 793 IC Sample Prep Module. Addi-

tionally, the suppressor block of the 753 Suppressor Module is inserted

and connected, which is described in detail in the corresponding

8.753.1001 Instructions for Use.

The connection of all parts should be done as described by the follow-

ing scheme:

B

Fill

Inject

A

SP

Module

A

752

793

709

MSM

753

B

Inject

Fill

A

SP

Module

A

752

793

709

MSM

753

B

Inject

Inject

A

SP

Module

A

752

793

709

MSM

753

H

2

O

Regenerant

Regenerant

Eluent

H

2

O

H

2

O

H

2

O

Regenerant

Regenerant

Eluent

H

2

O

H

2

O

H

2

O

Regenerant

Regenerant

Eluent

H

2

O

H

2

O

A) Sample loop at valve B

is filled.

B) Sample is neutralized

by SP Module A;

Anions are preconcen-

trated via valve A..

C) Preconcentrated

anions are eluted to

the separation column.

sample

sample

sample

preconcentration

column

separation

column

separation

column

separation

column

preconcentration

column

preconcentration