4.2 Maintenance and servicing

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

33

4.2.6 Cleaning the SP Module A

Cleaning the SP Module A may be necessary in the following cases

(see also section 4.1):

x Blockage of the SP Module A which cannot be remedied

(the SP Module A can no longer deliver solutions)

x Obstruction of the actuator which cannot be remedied

(the actuator can no longer be switched to next position)

To clean the connection piece and the rotor proceed as follows (see

Fig. 9):

1 Disconnect SP Module A from IC system

2 Dismantle SP Module A

x Unscrew nut 3

3337777 from actuator holder 40

40

40

40.

x Pull out connection piece 38

38

38

38 and actuator rotor 33339999 from

actuator holder (the connection piece and the rotor normally

stick together).

x Loosen connection piece from actuator rotor.

3 Clean input and output leads

x Each of the 6 capillary tubings attached to connection piece

are connected to the pump one after another and ultrapure

water pumped through.

x Check whether solution emerges from connection piece. If

one of the input or output leads remains blocked then the

connection piece must be replaced (order number

6.2835.010).

4 Clean actuator rotor

x Clean the sealing surface of actuator rotor 37

37

37

37 with the aid of a

lint-free cloth and ethanol.

5 Insert actuator rotor

x Insert actuator rotor in actuator holder in such a way that the

tubing connections at the rear of the rotor fit in the corre-

sponding openings inside the rotor and that one of the three

holes in the rotor can be seen from below in one of the open-

ings of the holder.

x If the rotor has been inserted correctly its sealing surface will

be about 4 mm inside the holder. If this is not the case then

the rotor must be brought into the correct position from below

with the aid of a sharp object (e.g. screwdriver).