2 Installation

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

8

2.2 Mains connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!



2.2.1 Setting the mains voltage

Before switching on the 793 IC Sample Preparation Module for the first

time, check that the mains voltage set on the instrument (see Fig. 2)

matches the local mains voltage. If this is not the case, you must reset

the mains voltage on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 19

19

19

19 of the

793 IC Sample Preparation Module.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 20

20

20

20 below the mains

connection 19

19

19

19 and take out completely.

3 Check fuse

Carefully take the fuse installed for the desired mains voltage

out of fuse holder 20

20

20

20 and check its specifications (the position

of the fuse in the fuse holder is marked by the white arrow

imprinted next to the mains voltage range):
100}

}

}

}120 V 0.5 A (slow-blow) Metrohm

No.

U.600.0013

220}

}

}

}240 V 0.25 A (slow-blow) Metrohm

No.

U.600.0010

4 Insert fuse

Change fuse if necessary and reinsert in fuse holder 20

20

20

20.

5 Install fuse holder

Depending on the desired mains voltage, insert fuse holder in

the 793 IC Sample Preparation Module so that the correspond-

ing mains voltage range can be read normally and the adjacent

white arrow points to the white bar imprinted below the fuse

holder (see Fig. 4).