2 Installation
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
8
2.2 Mains connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with the mains voltage set wrongly and/or
wrong mains fuse there is a danger of fire!
2.2.1 Setting the mains voltage
Before switching on the 793 IC Sample Preparation Module for the first
time, check that the mains voltage set on the instrument (see Fig. 2)
matches the local mains voltage. If this is not the case, you must reset
the mains voltage on the instrument as follows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 19
19
19
19 of the
793 IC Sample Preparation Module.
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 20
20
20
20 below the mains
connection 19
19
19
19 and take out completely.
3 Check fuse
Carefully take the fuse installed for the desired mains voltage
out of fuse holder 20
20
20
20 and check its specifications (the position
of the fuse in the fuse holder is marked by the white arrow
imprinted next to the mains voltage range):
100}
}
}
}120 V 0.5 A (slow-blow) Metrohm
No.
U.600.0013
220}
}
}
}240 V 0.25 A (slow-blow) Metrohm
No.
U.600.0010
4 Insert fuse
Change fuse if necessary and reinsert in fuse holder 20
20
20
20.
5 Install fuse holder
Depending on the desired mains voltage, insert fuse holder in
the 793 IC Sample Preparation Module so that the correspond-
ing mains voltage range can be read normally and the adjacent
white arrow points to the white bar imprinted below the fuse
holder (see Fig. 4).