2.2 Mains connection

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

9

13

13

13

13

220 ---- 240 V

100 ---- 120

V

19

19

19

19

20

20

20

20

220 ---- 240

V

100 ---- 120 V

Fig. 4: Setting the mains voltage

2.2.2 Fuses

One of the two fuses 0.5 A/slow-blow for 100}120 V or 0.25 A/slow-

blow for 220}240 V is installed in fuse holder 20

20

20

20 of the 793 IC Sample

Prep Module as standard.

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

For checking or changing fuses, proceed as described in section 2.2.1.

2.2.3 Mains cable and mains connection

Mains cable

The instrument is supplied with one of three mains cables:

x 6.2122.020 with plug SEV 12 (Switzerland, })
x 6.2122.040 with plug CEE(7), VII (Germany, })
x 6.2133.070 with plug NEMA 5-15 (USA, })

which are three-cored and fitted with a plug with an earthing pin. If a dif-

ferent plug has to be fitted, the yellow/green lead (IEC standard) must

be connected to protective earth (protection class I).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Mains connection

Plug the mains cable into mains connection plug 19

19

19

19 of the 793 IC

Sample Prep Module (see Fig. 4).

2.2.4 Switching the instrument on/off

The 793 IC Sample Prep Module is switched on and off using mains

switch 13

13

13

13. When the instrument is switched on lamp 1111 lights up.

100 – 120 V

220 – 240 V

13

13

13

13 Mains switch