3.3 Operation via 732 IC Detector

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

27

Rotate SP Module A at an event

In the 732 IC Detector the key can be used to

program a maximum of 4 different events (see section 4.7.3 of

732/733 Manual ). At each event the SP Module A can be brought

into the next position by setting remote output lead 2 to 1 (on,

active, 0 V). This lead must later be returned to the setting 0 (off,

inactive, open) before the SP Module A can be switched to the

next position.

Rotate:

EVENT YY-MM-DD HH:MM:SS
remote

1

✴✴✴✴✴✴

Reset

output lead:

EVENT YY-MM-DD HH:MM:SS
remote

0

✴✴✴✴✴✴

SP Module A position display

STEP

ACTUATOR

The position display 7

777 of the 793 IC Sample Prep Module beside

switch 8

888 shows whether the SP Module A is in the right position:

Lamp on

SP Module A in the right position

Lamp off

SP Module A in the wrong position or blocked