2.3 Connection to a Modular IC System for Neutralization

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

17

start-neutralization.smt

2. Save the system file “neutralization.smt” with a new name “start-

neutralization.smt“.

3. Remove the data recorder by clicking on the recorder icon with the

right mouse button an selecting 'unlink'.

4. Change this program for 'start-neutralization.smt' as follows and save

it again:

001 Ctrl INIT

732

002

Ctrl

PUMP 752 on

003 Wait 300

s

004

Ctrl

STEP MSM 753

005 Wait 300

s

006

Ctrl

STEP MSM 753

007 Wait 300

s

Initialize remote interface
Start 752 Pump Unit
Rinsing time
Switch SP Module A to next position
Rinsing time
Switch SP Module A to next position
Rinsing time

end-neutralization.smt

5. Save the program for “start-neutralization.smt“ under a new name

„end-neutralization.smt“.

6. Change the program for “end-neutralization.smt' as follows and save

it again:

001 Ctrl INIT

732

002

Ctrl

PUMP 752 off

Initialize remote interface
Stop 752 Pump Unit

7. The sample queue then contains the following entries, for example:

System Ident

Vial Chrom.-No.

1

Start-Neutralization.smt dummy 1

-

2

Neutralization.smt Probe

1

1

1

3

Neutralization.smt Probe

2

2

2

4

Neutralization.smt Probe

3

3

3

5

Neutralization.smt Probe

4

4

4

6

Neutralization.smt Probe

5

5

5

7

End-Neutralization.smt dummy 5

-

This example describes the processing of a queue with five samples.

The column 'Chrom. No.' is just given for information.