Metrohm 793 IC Sample Preparation Module User Manual
Page 21

2.3 Connection to a Modular IC System for Neutralization
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
17
start-neutralization.smt
2. Save the system file “neutralization.smt” with a new name “start-
neutralization.smt“.
3. Remove the data recorder by clicking on the recorder icon with the
right mouse button an selecting 'unlink'.
4. Change this program for 'start-neutralization.smt' as follows and save
it again:
001 Ctrl INIT
732
002
Ctrl
PUMP 752 on
003 Wait 300
s
004
Ctrl
STEP MSM 753
005 Wait 300
s
006
Ctrl
STEP MSM 753
007 Wait 300
s
Initialize remote interface
Start 752 Pump Unit
Rinsing time
Switch SP Module A to next position
Rinsing time
Switch SP Module A to next position
Rinsing time
end-neutralization.smt
5. Save the program for “start-neutralization.smt“ under a new name
„end-neutralization.smt“.
6. Change the program for “end-neutralization.smt' as follows and save
it again:
001 Ctrl INIT
732
002
Ctrl
PUMP 752 off
Initialize remote interface
Stop 752 Pump Unit
7. The sample queue then contains the following entries, for example:
System Ident
Vial Chrom.-No.
1
Start-Neutralization.smt dummy 1
-
2
Neutralization.smt Probe
1
1
1
3
Neutralization.smt Probe
2
2
2
4
Neutralization.smt Probe
3
3
3
5
Neutralization.smt Probe
4
4
4
6
Neutralization.smt Probe
5
5
5
7
End-Neutralization.smt dummy 5
-
This example describes the processing of a queue with five samples.
The column 'Chrom. No.' is just given for information.