2 Installation

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

22

2.4.3 Settings in the «IC Net 2.1» program

During configuration of the 793 IC Sample Prep Module, the IC Net

assistant “New system wizard” requires the definition of an “Event line”

for the SP Module A. Although this is not installed you must enter any

unused “Event line” to enable the program to proceed. Actually, the SP

Module A is controlled by the 766 sample processor (see Fig. 8).

The signal for switching the internal IC Suppressor Module is not

provided by the 766 program. For this purpose, in the IC Net program

the 733 IC Separation Center parameter “Auto Step” must be set to

“fill”.

The peristaltic pumps of the 752 IC Pump Unit and 793 IC Sample Prep

Module are started when the IC system is activated (measuring

baseline). For this , the option ’Start pump with startup hardware

Start pump with startup hardware

Start pump with startup hardware

Start pump with startup hardware’ has

to be selected for both devices.

catex.smt
Program for the 766 IC Sample Processor:

001 Ctrl INIT

732

002 Move

sample

003 Lift work

004 Ctrl ZERO

1

005

Ctrl

FILL A 1

006 Pump

120

s

007 Ctrl INJECT

A1

008

Ctrl

STEP MSM 753

Initialize remote interface
Move needle to sample position
Place lift with needle to working position
Trigger autozero at 732 IC Detector
Switch injection valve A at 733 to "Fill"
Pump sample to sample loop via SP Module A (*)
Switch injection valve A at 733 to "Inject"
Switch SP Module A to next position

(*) The time for pumping the sample to the SP Module A should be as small as

possible to minimize the loading of the cation exchange material.