4.2 Maintenance and servicing

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

35

4.2.7 Exchanging the cation exchange material

Exchanging the cation exchange material in the SP Module A of the 793

IC Sample Prep Module may be necessary in the following cases (see

also section 4.1):

x Loss of cation exchange capacity which cannot be remedied

x Blockage of the SP Module A which cannot be remedied

Both the 6.2835.000 actuator rotor as well as the 6.2835.010 connection

piece with the input and output leads can be exchanged. To exchange

these components proceed as follows (see Fig. 9):

1 Disconnect SP Module A from IC system

x Disconnect all input and output leads of the SP Module A

from IC system and the pump of the 793 IC Sample Prep

Module.

2 Dismantle actuator

x Unscrew nut 37

37

37

37 from actuator holder 40

40

40

40.

x Pull out connection piece 38

38

38

38 and actuator rotor 39

39

39

39 from

actuator holder (the connection piece and the rotor normally

stick together).

x Loosen connection piece from actuator rotor.

3 Clean actuator rotor

x Clean the sealing surface of new actuator rotor (6.2835.000)

with the aid of a lint-free cloth and ethanol.

4 Insert actuator rotor

x Insert new actuator rotor in actuator holder in such a way that

the tubing connections at the rear of the rotor fit in the corre-

sponding openings inside the rotor and that one of the three

holes in the rotor can be seen from below in one of the open-

ings of the holder.

x If the rotor has been inserted correctly its sealing surface will

be about 4 mm inside the holder. If this is not the case then

the rotor must be brought into the correct position from below

with the aid of a sharp object (e.g. screwdriver).

5 Clean connection piece

x Clean the sealing surface of new connection piece

(6.2835.010) with the aid of a lint-free cloth and ethanol.

6 Insert connection piece

x Insert new connection piece in actuator holder in such a way

that connection "1" is at the top and that the three lugs on the

connection piece fit in the corresponding openings of the

holder.

x Screw nut onto the thread of actuator holder manually (do not

use tools).