4.2 Maintenance and servicing

4.2.1 Care

The 793 IC Sample Prep Module requires proper care and attention.

Excessive contamination of the instrument could possibly lead to mal-

functions and a shorter service life of the inherently rugged mechanical

and electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the

interior of the instruments the mains plug of the 793 IC Sample Prep

Module must be immediately disconnected to prevent extensive

damage to the instrument electronics. Inform Metrohm service if your

instrument has been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in section 1.4.1.

4.2.2 Maintenance by Metrohm service

Maintenance of the 793 IC Sample Prep Module is best done as part of

an annual service performed by specialists from the Metrohm company.

Due to caustic and corrosive chemicals which are processed by this

device, it may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.2.3 Shutdown

If the IC system is shut down for a considerable length of time, the en-

tire IC system (without column, suppressor and SP Module A) must be

rinsed free from salt with methanol/water (1:4) to avoid crystallization

of eluent salts with the corresponding subsequent damage. In order to

rinse them the separating column, suppressor and actuator are discon-

nected and the injector and detector are directly connected to each

other. Rinse with methanol/water (1:4) until the conductivity drops below

10 PS/cm.
All three suppressor and actuator units are then rinsed for about 5 min

with ultrapure water.