beautypg.com

Metrohm 793 IC Sample Preparation Module User Manual

Page 16

background image

2 Installation

793 IC Sample Prep Module/ 8.793.1003 Instructions for Use

12

The inlet and outlet capillaries mounted on the SP Module A are then

connected up. Proceed as follows:

1 Insert SP Module A

x Place SP Module A 21

21

21

21 on the floor of the inner chamber of

the 733 Separation Center.

2 Connect SP Module A

x Remove the plastic stopper from opening 41

41

41

41 or 43

43

43

43 on the

733 IC Separation Center and push cable 23

23

23

23 mounted on the

SP Module A 21

21

21

21 through the opening.

x Connect cable 23

23

23

23 to connection 17

17

17

17 “Actuator“ of the 793 IC

Sample Prep Module (see Fig. 5).

3 Connect sample line to 793 SP Module A

x Screw inlet capillary 24

24

24

24 marked with "Sample in" at SP Module

A connection 22

22

22

22 (see Fig. 6) onto connection “5” of injection

valve B using a 6.2744.010 compression fitting.

4 Connect SP Module A to valve A (Preconcentration

column)

x Screw outlet capillary 29

29

29

29 marked with “Sample out” onto

connection “1” of injection valve A using a 6.2744.010 com-

pression fitting.

x Please note, that the preconcentration column at valve A has

to be rinsed in counterflow when switched from “Fill” to “In-

ject”.

1

3

2

24

24

24

24

27

27

27

27

28

28

28

28

29

29

29

29

25

25

25

25

26

26

26

26

Fig. 6:

Connections at SP Module A

24

24

24

24 Inlet capillary for sample

25

25

25

25 Inlet capillary for acid

26

26

26

26 Outlet capillary for acid

27

27

27

27 Outlet capillary for H

2

O

28

28

28

28 Inlet capillary for H

2

O

29

29

29

29 Outlet capillary for sample

H

2

O

Regenerant

Waste

Sample in

Sample out

Waste