4.1 Malfunctions and their rectification
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
29
Actuator moves to next
position but display lamp
7777 no longer lights up
x SP Module A dry
x SP Module A temporarily in the
wrong position
x Braking range not correctly
adjusted
x Rinse all 3 actuator units with solution for
at least 10 min.
x Switch instrument off and then on again.
Switch off SP Module A remote operation
with key 6666
8888
7777 lights up.
Repeat this procedure 2-3 times; wait at
least 10 s each time before the
key is operated.
If display lamp 7777 does not light up each
time the position is changed then the
braking range must be readjusted (see
below).
x Readjust braking range
(see section 4.2.8).
Actuator is not switched
to next position (display
lamp 7777 no longer lights up)
x Connection to SP Module A is
interrupted
x SP Module A dry
x Actuator temporarily blocked
x Actuator dirty
x Mechanical fault in Actuator
x Check cable connection from SP Module
A block to 793 Sample Prep Module.
x Rinse all 3 actuator units with solution for
at least 10 min.
x Switch instrument off and then on again.
Switch off SP Module A remote operation
with key 6666
8888
7777 lights up.
Repeat this procedure 2-3 times; wait at
least 10 s each time before the
key is operated.
If display lamp 7777 does not light up each
time the position is changed then the
braking range must be readjusted (see
below).
x Clean actuator (see section 4.2.6).
x Exchange actuator (see section 4.2.7).