Limited warranty – SATEC PM296 User Manual
Page 3
i
LIMITED WARRANTY
The manufacturer offers the customer an 24-month functional warranty on the
instrument for faulty workmanship or parts from date of dispatch from the
distributor. In all cases, this warranty is valid for 36 months from the date of
production. This warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument
malfunction. The manufacturer accepts no responsibility for the suitability of the
instrument to the application for which it was purchased.
Failure to install, set up or operate the instrument according to the instructions
herein will void the warranty.
Your instrument may be opened only by a duly authorized representative of the
manufacturer. The unit should only be opened in a fully anti-static environment.
Failure to do so may damage the electronic components and will void the
warranty.
NOTE
The greatest care has been taken to manufacture and calibrate your instrument.
However, these instructions do not cover all possible contingencies that may
arise during installation, operation or maintenance, and all details and variations
of this equipment are not covered by these instructions.
For additional information regarding installation, operation or maintenance of this
instrument, contact the manufacturer or your local representative or distributor.
IMPORTANT
Please read the instructions this manual before performing installation, and
take note of the following precautions:
1. Ensure that all incoming AC power and other power sources are turned
OFF
before performing any work on the instrument. Failure to do so may
result in serious or even fatal injury and/or equipment damage.
2. Before connecting the instrument to the power source, check the labels
on the side of the instrument to ensure that your instrument is equipped with
the appropriate power supply voltage, input voltages, currents, analog output
and communication protocol for your application.
3. Do not connect the instrument to a power source if it is damaged.
4. Do not expose the instrument to rain or moisture.
5. CLEANING: Use only a DRY cloth to clean the instrument.
6. The secondary of an external current transformer must never be allowed
to be open circuit when the primary is energized. An open circuit can cause
high voltages, possibly resulting in equipment damage, fire and even serious
or fatal injury. Ensure that the current transformer wiring is made through