13 pfc setup menu – SATEC C192PF8-RPR User Manual
Page 44

Chapter 4 Setup Menus
39
4.13 PFC Setup Menu
This menu allows you to configure the PFC setup options and to set the PFC
operation mode. Table 4-11 lists the PFC setup options, their code names and
applicable ranges. For more information on the PFC options and operation, refer
to Chapter 5, PFC Operation.
To select and view a setup option:
Press to activate the middle window
Use the up/down arrow keys to scroll to the desired option. The current
value for this option appears in the lower window.
To change the value of the selected option:
Press to make the lower window active.
Press the up/down arrow keys to scroll to the desired value.
Press to store the selected value, or press to quit the
setup menu.
Table 4-11 PFC Setup Options (
∗
default setting)
Code Parameter
Options
Description
OFF
∗
The PFC is OFF/Shut down the PFC
Aut.1
Self-adapting automatic mode. The
capacitor banks are connected and
disconnected progressively starting from
the first available bank.
Aut.2
Optimizing automatic mode. The capacitor
banks are connected and disconnected in
pre-optimized steps to reduce the number
of switching operations.
run
PFC operation mode
Hand
Safe manual operation mode. The banks
are connected and disconnected
progressively starting from the first
available bank when user commands are
entered from the front panel or through
communications (see Section 4.14, PFC
Manual Mode Menu).
PF
True power factor (based on true RMS
calculations) - PFC mode
PF.H01
Fundamental harmonic's power factor
(power factor displacement) - PFC mode
rE.P
∗
True reactive power (based on true RMS
calculations) - RPR mode
triG
PFC setpoint trigger
rE.H01
Fundamental harmonic's reactive power -
RPR mode
SEtP
Operational setpoints
SP1
The target setpoint range is defined by
setpoint SP1
SELECT
SELECT
PFC
Aut.1
run
SELECT
CHG
ENTER
PFC
ENTER
ENTER
ESC