Heidolph LABOROTA 20 automatic User Manual
Page 98
98
E
Limit value:
Depending upon the desired vacuum, at which distillation should occur, it can be advantageous to
adjust the limit value. From the factory, the limit value is set at 900 mbar. The limit value can be
changed from 0 to 1200 mbar.
Pay attention that the limit value is never set lower than the value, at which
distillation should occur, as an „Emergency“ will be triggered otherwise.
Change limit value:
Switch off the LABOROTA 20 control with the master
switch „Power“.
Simultaneously press the Select key and the Start Stop
key in the „Speed“ field and switch on the master
switch „Power“. Wait until all LED´s light up and then
See also other documents in the category Heidolph Equipment:
- Hei-VAP Precision (59 pages)
- Hei-VAP Value (84 pages)
- Rotacool Mini (104 pages)
- Rotacool (15 pages)
- RotaChill (37 pages)
- ROTAVAC valve control (188 pages)
- ROTAVAC valve control (174 pages)
- ROTAVAC vario control (34 pages)
- ROTAVAC Vario pumping unit (61 pages)
- LABOROTA 4000 eco (168 pages)
- LABOROTA 4000 up to 4003 (364 pages)
- Heating bath HB 4000 (52 pages)
- LABOROTA 20 eco (180 pages)
- LABOROTA 20 compact and safety (284 pages)
- ROTAVAC 20 (35 pages)
- Automatic Module Distimatic (52 pages)
- Synthesis 1 (268 pages)
- MR Hei-Standard (39 pages)
- MR Hei-End (124 pages)
- MR Hei-Mix L (24 pages)
- MR Hei-Mix S (34 pages)
- EKT Hei-Con (78 pages)
- Heat-On (13 pages)
- StarFish Multi-Experiment Work Station (17 pages)
- RZR 1 (44 pages)
- RZR 2020 (58 pages)
- RZR 2051 control (68 pages)
- REAX top (44 pages)
- REAX 2 (48 pages)
- REAX 20 (54 pages)
- Multi Reax (60 pages)
- TITRAMAX 100 / 101 / 1000 (60 pages)
- UNIMAX 1010 (68 pages)
- UNIMAX 2010 (72 pages)
- Incubator 1000 (64 pages)
- Pump drives (198 pages)
- SilentCrusher M (146 pages)
- SilentCrusher S (98 pages)