E

Reaching this final pressure during distillation, will discontinue distillation right away. Also refer to

Auto Start Stop function (see page 94)

If distillation should not be discontinued, enter "1" in "Endp mode", which never can
be reached by physical condition.

For hysteresis setting select LED

Vapor. Use Set keys as required

20. Calibrate Vacuum pressure sensor

Install separate calibrated vacuum gauge in vacuum

line close to condenser

Unplug power plug of automatic module from socket

of base unit

Unplug power plug of vacuum pump from socket of

the backside of automatic module and plug into plug
on the backside of base unit