Turn vacuum pump ON

Turn LABOROTA 20 control automatic Master switch (Power)

OFF

Depress Select and Start Stop keys next to display I together and hold for 5 seconds, turn

master switch ON, release keys

Display Speed shows “CAL”, display Time “VAC”.

Display I to show 750 and display II 1 (1

st

step)

Increase/decrease separate vacuum gauge reading of in display I with Set key

Confirm with Start Stop key

Display I

to show “50” and display II “2” (2

nd

step).

LABOROTA 20 control automatic automatically is going to turn its vacuum pump ON, and

continues evacuating to max. limit (limit defined by vacuum
pump's power). To this end disconnect aerating valve
control wire. (Vacuum gauge display won't change any
more).

Increase/decrease vacuum gauge reading in Display I

with Set key

Confirm with Start Stop key located next to Display I. As

confirmation for the accepted calibration values, „0“
appears in Display I and Display II

Turn main switch OFF and ON

Connect aerating valve control wire again

750

1

Display I

Select key

Taste

Start /Stop

Display II

Display Speed

Display Time

CAL

VAC

increase

decrease