CONTROL PANEL

13. Master switch / EMERGENCY-CUT-OUT


The Master switch disconnects all wires from power supply, and serves as a RESET button.
(after Emergency CUT-OUT).

The Emergency CUT-OUT discontinues operation of the rotary evaporator immediately.

Stop rotation

Lowers heating bath

Opens aerating valve (vacuum valve closes)

Turns OFF heating circuit

Product feed and condensate discharge is cancelled

Buzzer sounds

Emergency LED turns ON

CAUTION: if entry of ambient air causes unwanted reactions, take adequate action
(inert gas, or keep aerating valve closed).

System re-starts after an Emergency CUT-OUT situation requires system RESET. Make sure the
problem is fixed before resetting. See an expert, if required.

To Reset, turn Master switch OFF & ON.

14. Heating bath lift

Heating bath lift RAISES, as long as the Lift UP
key is held depressed.
Heating bath lift LOWERS, as long as the Lift DOWN
key is held depressed.
At upper and lower limits, lift motion stops automatically.

Water bath only:
The LABOROTA 20 control automatic WB (water bath)
features a fully automatic water refill system with float switch.

Lift UP

Master switch

Emergency CUT-OUT

Lift
DOWN

Stop-Mode

Master switch

Emergency CUT-OUT

Stop-Mode