beautypg.com

Heidolph LABOROTA 20 automatic User Manual

Page 112

background image

112

E

41. RS 232 interface commands

query

Command

action

remarks

Act commands

VAC?

Act Vacuum

VAP?

Act Vapor

BATH?

Act Bath

CHILL?

Act Chiller

SPEED?

Act Speed

TIME?

Act Time

elapsed time, continuous operation

COUNT?

Act Time

time, timer mode

Safety functions

SC?

Safety Cover

0 = action; 1 = no action

FS?

Flask Support

0 = action; 1 = no action

RS?

Rotation Stop

0 = action; 1 = no action

OP?

Over-
Temperature
Prot.

0 = action; 1 = no action

triggering

Command

action

remarks

Set commands

VAC=XXXX!

Set Vacuum

4-digit in mbar; range set 0000 to 1200

BATH=XXXX!

Set Bath

4-digit, in 1°C increments
range set 0000 to 1800 equals 0.0 to 180.0

SPEED=XXX!

Set Speed

3-digit, in rpm; range set 000 to 180

TIME=XXXX!

Set Timer

4-digit, in hh:mm
range set 0000 to 9959 equals 00:00 to 99:59

Actions

SVX!

Start Stop Vacuum

SV0! = OFF ; SV1! = ON

VVX!

Valve

vacuum valve; VV0! = CLOSE ; VV1! = OPEN

BVX!

Ventilation

aerating valve; BV0! = CLOSE ; BV1! = OPEN

SBX!

Start Stop Bath

SB0! = heater OFF ; SB1! = heater ON

SSX!

Start Stop Speed

SS0! = rotation OFF ; SS1! = rotation ON

LU!

lift UP

LD!

lift DOWN