beautypg.com

11 technical data, 1 power supply, 2 ambient conditions – Eppendorf D30 BioPhotometer User Manual

Page 69: 3 weight/dimensions, Power supply, Ambient conditions, Weight/dimensions

background image

69

Technical data

Eppendorf BioPhotometer

®

D30

English (EN)

11

Technical data

11.1

Power supply

11.2

Ambient conditions

Do not expose to direct sunlight.

11.3

Weight/dimensions

Power supply

100 V to 240 V ±10 %, 50 Hz to 60 Hz

Overvoltage category

II

Degree of pollution

2

Power consumption

Maximum power consumption according to name plate: 25 W
Approx. 15 W during operation
Approx. 5 W with the display dimmed

Permitted mains interruption

Approx. 10 ms at 90 V
Approx. 20 ms at 230 V

Protection class

I

Fuses

T 2.5 A/250 V, 5 mm × 20 mm (2 pcs.)

Operation

Ambient temperature: 15°C to 35°C
Rel. humidity: 25% to 70%
Air pressure: 86 kPa to 106 kPa

Air pressure

Use up to an altitude of 2000 m above MSL

Weight

5.4 kg

Dimensions

Width: 295 mm
Depth: 400 mm
Height:150 mm

Space required

Width: 500 mm (with thermal printer: 750 mm)
Depth: 500 mm