Installation
Eppendorf BioPhotometer
®
D30
English (EN)
4
Installation
4.1
Preparing installation
Keep the transport carton and the packing material for subsequent safe transport or storage.
Check the completeness of the delivery using the information in the delivery package (see Delivery
Check all parts for any transport damage.
4.2
Selecting the location
Select the location for the BioPhotometer D30 according to the following criteria:
• 2 grounded sockets for the BioPhotometer D30 and for the printer.
• Solid laboratory bench with horizontal work surface
Space requirement of the device: 50 cm (with printer: 75 cm) width, 50 cm depth.
• Temperature: 15°C to 35°C.
• Avoid temperature fluctuations (e.g, caused by open windows).
• Avoid direct sunlight.
• Humidity: 25% to 70% relative humidity.
4.3
Connecting the device to the mains
1. Place the BioPhotometer D30 on a suitable work surface.
2. Verify that the mains/power supply voltage and mains/power frequency match the information on the
name plate.
3. Connect the device to the mains/power line and switch it on with the power switch.
4. Remove the protective film from the display.
Ensure that no objects (e.g., loose sheets, notebooks) that could impede the flow of air are
positioned under the device.