Eppendorf D30 BioPhotometer User Manual
Page 66
![background image](/manuals/551302/66/background.png)
Troubleshooting
Eppendorf BioPhotometer
®
D30
English (EN)
66
The coefficient of
determination for
the regression
evaluation of the
standard series is <
0.8.
• For methods with evaluation of
standard series via the regression
procedure: If the regression
evaluation for the standard series was
nonlinear, but the standard evaluation
was accepted by the user, a warning
appears after samples have been
measured.
Use the sample results with the
reservation mentioned or repeat the
measurement of the standard series
and samples.
Scan: Some of the
measured
absorbances are
too high and are
not displayed.
• For at least one scan wavelength, the
absorbance exceeded the measuring
range.
• Turbidity of the measuring solution.
• Measurements at the limits of the
photometric measuring range.
If the non-displayed areas of the scan
are relevant: Dilute the sample or
remove the dilution via centrifugation
and repeat the measurement.
Absorbance at the
measuring
wavelength is too
high.
• Turbidity of the measuring solution.
• Optical surfaces of the cuvette are
soiled.
• Cuvette has been inserted into the
cuvette shaft facing the wrong
direction.
• Too high absorbance of measuring
solution.
Measure again considering the
possible causes.
The calculated
result is negative.
• Measuring solution not prepared
correctly.
• The incorrect factor has been entered
(wrong algebraic sign).
Measure again considering the
possible causes.
The result has more
than 6 pre-decimal
places.
• Very high sample concentration.
• Concentration unit does not match
the expected range of the sample
concentrations.
Dilute sample and measure again.
Change the concentration unit
(Parameter
Unit) and measure again.
The result is more
than 5 % outside of
the standard
concentration
range.
• For methods with evaluation via
standard curves (nonlinear evaluation
method):
The sample result is more than 5 %
outside of the standard concentration
range.
Remeasure the sample under
conditions under which the result is
within the range of the standard
concentrations (dilute sample, modify
standard concentrations and
remeasure).
• Calculation not
possible
because of
division by zero.
Absorbance
result is zero.
• Calculation
error. Division
by zero.
• The evaluation required dividing by
an absorbance result with the value of
"zero". This is not mathematically
permissible.
Examples: Calculation of a factor at
one-point calibration; calculation of a
260/280 ratio with nucleic acid
measurements.
Check the reagents and samples used
and repeat the measurement.
Problem
Cause
Solution