8.3.1.1 Checking the photometric accuracy

1. In the

Device calibration group, select the

Photometer unit function and confirm with
enter.

2. Select if you want to check the wavelength

systematic error or the photometric accuracy.
Confirm with

enter. Press [Next >] to switch to

the measurement.

3. Follow the instructions on the display and first

measure the blank filter A0 and then the first test
filter A1.
The device will measure the test filter 15 times at
6 wavelengths and then show the average values
as well as the variation coefficients of the series
of measurement for all wavelengths.