beautypg.com

3 method parameters, Method parameters – Eppendorf D30 BioPhotometer User Manual

Page 32

background image

Methods
Eppendorf BioPhotometer

®

D30

English (EN)

32

6.3

Method parameters

This chapter illustrates the parameters for programming the methods. The order of the parameters in the
device display may slightly differ from the order in the table in order to display the parameters more clearly.
The table displays all parameters available for the various methods. Only a small portion of these
parameters are required for the corresponding method and will be shown in the display.

Parameter

Entry

Explanation

Cuvette

Selection:
10 | 5 | 2 | 1 | 0.5 | 0.2 |
0.1 mm

Optical path length of the cuvette. The device always
automatically converts absorbance values to the 10 mm path
length of a standard cuvette (see Absorbance values on p. 73).
Therefore, there is no need to change factors such as "50" for
the calculation of dsDNA concentrations when modifying the
Cuvette parameter.

Wavelength

Selection:
230 | 260 | 280 | 320 |
340 | 405 | 490 | 562 |
595 | 600 nm

Measurement wavelength: The concentration is calculated
based on the absorbance measured with this wavelength.
For some method groups (e.g.,

Nucleic acids and Proteins

direct UV), the wavelengths are preprogrammed.

Unit

Selection:
mg/mL | μg/mL | ng/
mL | pg/mL | μg/μL |
mg/dL | μmol/mL |
nmol/mL | pmol/mL |
pmol/μL | U | U/mL |
U/L | % | Abs | A/min
In addition, further
units are freely
programmable in the
General Method
Parameters/Units
function. Max. 7 digits.

Unit for the concentration result.
In the preprogrammed methods of the Routine group, the
selection is restricted to units that are useful for these methods.

Calculation

Selection:
factor | standard

Evaluation procedure for the calculation of the sample
concentration from the measured absorbance.

Factor

Value input:
Factor.
Limit: max. of 6 digits
including decimal
point.

Factor for converting absorbance values into the concentration.
For the

Factor method group, you also can enter negative

factors.

Protein

Selection:
List of protein types
which are stored in the
General Method
Parameters/Proteins
function.

Only for the

Proteins direct UV method group.

When selecting the protein, the corresponding

Factor

parameter programmed in the

General Method Parameters/

Proteins function also will be imported from that function.