8.3.2

Device self test

You can set the frequency of the automatic self test (taking approx. 1 minute) with the

Device settings

function (see Device settings on p. 51). Ex-factory, the

self test interval is set to "weekly".

The self test checks the following items:

• Verification of the detector

– Determination of the random error of the available wavelengths

• Verification of the light source

– The device checks the maximally available light source energy and the light cable quality
– Determination of the random error of a signal on the reference sensor
– Determination of the signal level at the reference sensor
– Separate determination of the light intensity in the UV range

• Determination of the systematic and random error of the wavelength

 In the Device calibration group, select the Perform selftest function and confirm with enter.

After the self test has passed, the display shows the message

PASSED.

If the display shows the message

FAILED, the self test has failed. If this error cannot be corrected (see

Error messages on p. 62), contact Eppendorf Service.

8.4

Replacing fuses

The fuse holder is located between the mains connection socket and the mains power switch.

1. Disconnect the power plug.

2. Press the upper and lower end of the plastic springs

1 together and pull the fuse holder 2 fully out.

3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail

3 is positioned

correctly.

DANGER! Electric shock.

 Switch off the device and disconnect the power plug before starting maintenance or

cleaning work.