13 Testing the photometer

To enable the photometric accuracy and the wavelength accuracy to be tested, a filter set (secondary UV-VIS
filter) is available from Eppendorf. This set contains three filters ("Sample A

1

", "Sample A

2

" and "Sample A

3

")

for testing the photometric accuracy and two filters ("Sample 260 nm" and "Sample 280 nm") for testing the
wavelength accuracy. The absorbance of the filters is measured against a blank filter ("Blank A

0

").

To carry out these measurements, blank filters and "sample filters" (test filters) are inserted into the cuvette
holder in the same manner as cuvettes. When doing so, please ensure that the label with the filter description
is facing the user. The absorbance values measured for the test filters are compared to those within the range
of permitted values. The limits for the permitted range are contained in a table found on the inside of the lid of
the filter box (see Figure: "X.XXX – X.XXX A").

Fig.: Inside of the lid of the filter box

BioPhotometer

Function: PHOTOMETERTEST

Secondary / Sekundär - UV - VIS - Filter

Order No. / Best.Nr.: 6131 928.007

Limits
Grenzwerte

measured against Blank A

0

at approx. 20

°C

gemessen gegen Blank A

0

bei ca. 20

°C

6131

914.XXX

916.XXX

917.XXX

921.XXX

922.XXX

923.XXX

Filter

Type

Blank

A

0

Sample

260 nm

Sample

280 nm

Sample

A

1

Sample

A

2

Sample

A

3

Reference

Systematic error / Systematische Messabweichung

empty

leer

of wavelength / der Wellenlänge

of photometer / des Photometers

Traceable to / rückführbar auf

Nist: SRM 2034, SN: 99

Nist: SRM 2031a, SN: 577

(A) / (E)

Limiting values (A) / Grenzwerte (E)

230 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

260 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

280 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

320 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

562 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

595 nm

0.000

X.XXX - X.XXX

X.XXX - X.XXX

X.XXX - X.XXX

230 –

595 nm

Random error / Zufällige Messabweichung

of wavelength / der Wellenlänge

of photometer / des Photometers

Limiting values CV (%) / Grenzwerte VK (%)

0.000

≤ 3.0 %

≤ 3.0 %

≤ 1.0 %

≤ 1.5 %

Please protect against dust, heat and liquid

The limits are valid for max. 2 years.

Bitte vor Staub, Hitze und Flüssigkeiten schützen

Die Grenzwerte gelten für max. 2 Jahre.

Date

Signature

Datum

Unterschrift