beautypg.com

6 programming, 3 explanation of parameters – Eppendorf BioPhotometer User Manual

Page 30

background image

72

6 Programming

6.3 Explanation of parameters

Parameters are defined as selection parameters or as parameters for entering numbers. In the case of
selection parameters, the programmable alternatives are method-dependent (see overview in previous
section).

Parameter

Entries

Explanation

Calculation

Selection

Selection of calculation procedures: Absorbance, Factor,
Standard and Warburg formula.
In the case of calculation using the Warburg formula, the
measured value for A

260

is marked in the results display

and on the results printout with a "

".

Factor

Entry of numbers
(five-figure)

(Only when the calculation process "Factor" has been
selected)
Entering a factor;
the number of decimal places determines the number of
decimal places in the result.

Corr. with A

320

Selection

(Only for nucleic acid methods and for the direct photometric
determination of protein)
Selection from "Corr. with A

320

off" and "Corr. with A

320

on";

"Corr. on" means: the absorbance measured at 320 nm is
subtracted from the absorbance results at 260, 280 and
230 nm. Example of application: Correction of turbidity in the
sample.
When the correction function is switched on, the measuring
value for A

320

is marked with a "

" in the results display and

on the results printout.

Unit

Selection

The selection from preprogrammed concentration units is
method-dependent.

M. unit
(molar unit)

Selection

Selection is method-dependent
(for nucleic acid measurements only);
is required for the conversion of the concentration into molar
concentrations (

key).

Cuvette

Selection

Selection from 10 mm, 5 mm, 2 mm and 1 mm optical path
length; the result is converted for an optical path length of
10 mm (see Section 12 "Calculation").

Conversion

06_Programm_e.fm Seite 72 Montag, 20. Februar 2006 10:54 Uhr