8 error messages, result flagging and help texts, Result flagging error texts in the results display – Eppendorf BioPhotometer User Manual
Page 35

77
8 Error messages, result flagging and help texts
Result flagging
Error texts in the results display
Flagging
Explanation
1.586 A
260
Flagging of A
260
in the display or on the printout
(for method "Protein direct" only):
The method was calculated using the Warburg formula.
0.015 A
320
Flagging of A
320
in the display or on the printout
(for method "Protein direct" and for nucleic acid methods only):
The absorbances at 260, 280 and 230 nm are corrected with the absorbance
at 320 nm (see Section 6, "Programming").
Error text
Explanation / Cause
Solution
+++++
The absorbance measured is greater
than 3.0 A.
– Dilute the sample.
– Check the cuvette
(height of light path must
be 8.5 mm).
– Clean the cuvette shaft
(see Section 9).
– Insert the cuvette correctly (the
measuring window must be facing
the light path).
– Use a cuvette made of material
that transmits light at the
measuring wavelengths used
(e.g. quartz glass or UVette
®
from
Eppendorf for nucleic acid
measurement).
!
!
!
!
!
The calculated result cannot be displayed
(value too high).
Check the parameter
(Is the factor too high?).
– – – – –
(Instead of a value for the ratio:)
Ratio cannot be calculated because one of
the absorbance values used for calculating
the ratio is 0 A or > 3.0 A.
Repeat the measurement
(dilute sample if necessary).
▲
▲