8 Error messages, result flagging and help texts

Result flagging

Error texts in the results display

Flagging

Explanation

1.586 A

260

Flagging of A

260

in the display or on the printout

(for method "Protein direct" only):
The method was calculated using the Warburg formula.

0.015 A

320

Flagging of A

320

in the display or on the printout

(for method "Protein direct" and for nucleic acid methods only):
The absorbances at 260, 280 and 230 nm are corrected with the absorbance
at 320 nm (see Section 6, "Programming").

Error text

Explanation / Cause

Solution

+++++

The absorbance measured is greater
than 3.0 A.

– Dilute the sample.

– Check the cuvette

(height of light path must
be 8.5 mm).

– Clean the cuvette shaft

(see Section 9).

– Insert the cuvette correctly (the

measuring window must be facing
the light path).

– Use a cuvette made of material

that transmits light at the
measuring wavelengths used
(e.g. quartz glass or UVette

®

from

Eppendorf for nucleic acid
measurement).

!

!

!

!

!

The calculated result cannot be displayed
(value too high).

Check the parameter
(Is the factor too high?).

– – – – –

(Instead of a value for the ratio:)
Ratio cannot be calculated because one of
the absorbance values used for calculating
the ratio is 0 A or > 3.0 A.

Repeat the measurement
(dilute sample if necessary).