Error messages – RADWAG WTB 600/AU Carat Balance User Manual
Page 62
22. ERROR MESSAGES
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Adjustment mass or start mass beyond the acceptable
range (
±1% for weight, ±10 for start mass)
Err8
- Tarring / Zeroing operation time exceeded
Err10
- Accepting zero mass indication.
Accepting mass measured in liquid (water), that is above /
equal to mass weighed in the air
Err11
- Density of subsidiary metals (determined by the software)
is higher than density of the main metal. Refers to Mode A
(3-component alloy)
LO
- Caratage of pure gold is determined below the value
of 9.00K (37,50%). Refers to Mode G.
Density of platinum alloy is lower than density of subsidiary
metals. Refers to Mode P.
Determined density of metal alloy is below declared density
values of subsidiary metals d1, d2. Refers to Mode A
(2-component alloy)
HI
- Determined density of metal alloy is above the declared
density values of subsidiary metals d1, d2. Refers to Mode A
(2-component alloy)
null
- Zero value from the AD converter
FULL
2
- Measurement range (Max. capacity) exceeded
LH
- Start mass error, the mass on the weighing pan assembly
is beyond the acceptable range (-5% to +15% of start mass)
Caution:
1. Errors: Err2, Err3, Err4, Err5, Err8, Err10, Err11, LO, HI, null, that
appear on the display are also accompanied by a short beep sound
(about 1 sec.);
2. Error FULL2 that appears on the display is also accompanied by
a continuous sound until the cause of error disappears.
- 62 -