22. ERROR MESSAGES


Err2

- Value beyond the zero range

Err3

- Value beyond the tare range

Err4

- Adjustment mass or start mass beyond the acceptable

range (

±1% for weight, ±10 for start mass)

Err8

- Tarring / Zeroing operation time exceeded

Err10

- Accepting zero mass indication.

Accepting mass measured in liquid (water), that is above /
equal to mass weighed in the air

Err11

- Density of subsidiary metals (determined by the software)

is higher than density of the main metal. Refers to Mode A
(3-component alloy)

LO

- Caratage of pure gold is determined below the value

of 9.00K (37,50%). Refers to Mode G.

Density of platinum alloy is lower than density of subsidiary
metals. Refers to Mode P.

Determined density of metal alloy is below declared density
values of subsidiary metals d1, d2. Refers to Mode A
(2-component alloy)

HI

- Determined density of metal alloy is above the declared

density values of subsidiary metals d1, d2. Refers to Mode A
(2-component alloy)

null

- Zero value from the AD converter

FULL
2

- Measurement range (Max. capacity) exceeded

LH

- Start mass error, the mass on the weighing pan assembly

is beyond the acceptable range (-5% to +15% of start mass)


Caution:

1. Errors: Err2, Err3, Err4, Err5, Err8, Err10, Err11, LO, HI, null, that

appear on the display are also accompanied by a short beep sound
(about 1 sec.);

2. Error FULL2 that appears on the display is also accompanied by

a continuous sound until the cause of error disappears.