User adjustment – RADWAG WTB 600/AU Carat Balance User Manual
Page 47
• After the indication returns to ZERO place the tested sample in the
container with liquid (water) (weighing in liquid, flickering command
IN WATER) and on stabilization of measurement result press
key,
• Simultaneously, for approximately 2 seconds the software displays
liquid’s temperature value, and a window on process completion
appears (message END):
• The following test results are displayed on pressing
key.
Test results:
• Density of tested sample [g/cm
3
],
• Volume of tested sample [cm
3
],
• Liquid
temperature
[
°C],
• Mass of weighed sample in the air [g],
• Mass of weighed sample in liquid (water) [g].
At the same time the test results can be printed on a connected printer
by pressing button
.
17. USER ADJUSTMENT
Only for non-verified scales
Confirmation of high weighing accuracy requires periodical correcting
of adjustment factor stored in the balance’s memory – this process is
known as balance adjustment. Adjustment should be carried out on
balance start of operation or if dynamic change of temperature occurs.
Before carrying out adjustment procedure remove load from the weighing
pan.
- 47 -