• After the indication returns to ZERO place the tested sample in the

container with liquid (water) (weighing in liquid, flickering command

IN WATER) and on stabilization of measurement result press

key,

• Simultaneously, for approximately 2 seconds the software displays

liquid’s temperature value, and a window on process completion
appears (message END):

• The following test results are displayed on pressing

key.


Test results:

• Density of tested sample [g/cm

3

],

• Volume of tested sample [cm

3

],

• Liquid

temperature

[

°C],

• Mass of weighed sample in the air [g],
• Mass of weighed sample in liquid (water) [g].


At the same time the test results can be printed on a connected printer

by pressing button

.


17. USER ADJUSTMENT

Only for non-verified scales

Confirmation of high weighing accuracy requires periodical correcting
of adjustment factor stored in the balance’s memory – this process is
known as balance adjustment. Adjustment should be carried out on
balance start of operation or if dynamic change of temperature occurs.
Before carrying out adjustment procedure remove load from the weighing
pan.