A set of commands for rs interfaces – RADWAG WTB 600/AU Carat Balance User Manual
Page 54
21.2. A set of commands for RS interfaces
Commands
Description of commands
Z
Zero balance
T
Tare balance
TO
Give tare value
S
Send stable measurement result in basic measuring unit
SI
Immediately send measurement result in basic measuring unit
SU
Send stable measurement result in current measuring unit
SUI
Immediately send measurement result in current measuring unit
C1
Switch on continuous transmission in basic measuring unit
C0
Switch off continuous transmission in basic measuring unit
CU1
Switch on continuous transmission in current measuring unit
CU0
Switch off continuous transmission in current measuring unit
PC
Send all implemented commands
Caution:
1. Each command must end with CR LF characters;
2. The best policy for communication is not sending another
command until the former answer has been received.
21.3. Responses format for commands sent from computer level
On receipt of a command, the terminal sends a response in one of the
following formats:
XX_A CR LF
command understood and in progress
XX_D CR LF
command carried out (appears only after the command XX_A)
XX_I CR LF
command understood but not accessible at this moment
XX _ ^ CR LF
command understood but max range is exceeded
XX _ v CR LF
command understood but min range is exceeded
XX _ OK CR LF Command carried out (completed)
ES_CR LF
Command not recognized
- 54 -