Cooperation with printer – RADWAG WTB 600/AU Carat Balance User Manual
Page 51
Caution:
If the start mass adjustment lasts longer than 15 seconds lasts longer than
15 seconds, error message
emitted. Press
key repeat adjustment procedure under more stable
ambient conditions!
18. COOPERATION WITH PRINTER
Each time the
key is pressed a current mass value together with
measuring units is sent to RS 232 interface.
Depending on setting of STAB parameter, the printout is released with
stable or unstable (temporary) value. Depending on setting of REPL
parameter, printout is automatic or manual.
Balance WTB/AU enables cooperating with one of the following printers
KAFKA series:
a) KAFKA
Printout of weighing result and measuring unit only.
b) KAFKA
1/Z
This printer features an internal real time clock.
In addition to mass and measuring unit, both date and time can
be printed.
c) KAFKA SQ S
This printer features an internal real time clock and possibility of
running statistics from measurements. Statistic contents:
quantity of samples, total mass of weighed samples, average
value, standard deviation, variation factor, min value, max value,
difference max – min.
- 51 -