Caution:

If the start mass adjustment lasts longer than 15 seconds lasts longer than
15 seconds, error message
is displayed and short audible signal is

emitted. Press

key repeat adjustment procedure under more stable

ambient conditions!



18. COOPERATION WITH PRINTER

Each time the

key is pressed a current mass value together with

measuring units is sent to RS 232 interface.

Depending on setting of STAB parameter, the printout is released with
stable or unstable (temporary) value. Depending on setting of REPL
parameter, printout is automatic or manual.

Balance WTB/AU enables cooperating with one of the following printers
KAFKA series:

a) KAFKA

Printout of weighing result and measuring unit only.

b) KAFKA

1/Z

This printer features an internal real time clock.
In addition to mass and measuring unit, both date and time can
be printed.

c) KAFKA SQ S

This printer features an internal real time clock and possibility of
running statistics from measurements. Statistic contents:
quantity of samples, total mass of weighed samples, average
value, standard deviation, variation factor, min value, max value,
difference max – min.