Send measurement result in current measuring unit, Immediately – RADWAG WTB 600/AU Carat Balance User Manual
Page 58
Example:
S U CR LF – command from a computer
S U _ A CR LF – command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.
21.4.7. Immediately send measurement result in current measuring
unit
Format: SUI CR LF
Accessible responses:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- mass value in current measuring unit is returned immediately
Frame format:
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I
stability
marker
space sign mass space
unit
CR LF
Example:
S U I CR LF – command from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out,
immediate response of mass value in current measuring unit
Where: _ - space
21.4.8. Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Accessible responses:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit