Continuous transmission – RADWAG WTB 600/AU Carat Balance User Manual
Page 61
Example 1:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF –
a printout generated from a balance
on pressing ENTER/PRINT key.
Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF a printout generated from a balance
on pressing ENTER/PRINT key.
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - a printout generated from a balance
on pressing ENTER/PRINT key.
21.6. Continuous transmission
The balance enables operating in continuous transmission mode.
The mode is enabled or disabled in balance parameters or using RS232
commands.
The frame format sent by the balance in case of setting
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space
stability
marker
space sign mass space Unit CR
LF
Stability marker
[space] if stable
[?] if unstable
[^] if error of exceeding range to “+” occurs
[v] if error of exceeding range to “-“ occurs
sign
[space] for positive values or
[-] for negative values
mass
9 characters with decimal point and right justification
unit
3 characters with decimal point and left justification
command
3 characters with left justification
The frame format sent by the balance in case of setting
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I
stability
marker
space sign mass space
unit
CR
LF
- 61 -