Frame format, Format: c0 cr lf, Accessible responses – RADWAG WTB 600/AU Carat Balance User Manual
Page 59: Format: cu1 cr lf, Format: cu0 cr lf
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space
stability
marker
space sign mass space
unit
CR LF
21.4.9. Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and in progress
21.4.10. Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I
stability
marker
space sign mass space
unit
CR LF
21.4.11. Switch off continuous transmission in current measuring unit
Format: CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and in progress
- 59 -