Minimal mass threshold – RADWAG WTB 600/AU Carat Balance User Manual
Page 21
Pr_n noStAb
-
immediate printout
(not accessible in verified balances)
Pr_n StAb
-
sending stable measurement result
Pr_n rEPL
-
automatic operation
Pr_n CntA
-
continuous transmission in basic measuring
unit
Pr_n Cntb
-
continuous transmission in current measuring
unit
Return to weighing:
see 11.2.2.
13.2. Minimal mass threshold
This function is necessary while working with automatic tare or automatic
operation mode.
Automatic tare will not be applied until the indication (gross value) is lower
than the value set in S_Lo parameter.
In automatic operation measurements (net value) are sent via RS232
when the indication is equal or greater than the value set in S_Lo
parameter.
Procedure:
• Enter the submenu
- 21 -