Testing carat content of gold alloy, Mode’s local settings, Table of gold alloy density – RADWAG WTB 600/AU Carat Balance User Manual
Page 39: Density, Table of gold alloy

Test results:
• Gold
caratage
[K],
• Percent content of gold in tested sample [%],
• Density of tested sample [g/cm
3
],
• Volume of tested sample [cm
3
],
• Liquid
temperature
[
°C],
• Mass of weighed sample in the air [g],
• Mass of weighed sample in liquid (water) [g].
At the same time the test results can be printed on a connected printer
by pressing button
.
16.7. Testing carat content of gold alloy
Working mode MODE K enables testing one of three gold alloys:
• Gold with admixture of silver and copper (mode K-1),
• Gold with admixture of copper (mode K-2),
• Gold with admixture of silver (mode K-3).
16.7.1. Mode’s local settings
Local settings of MODE K are grouped in submenu
No.
Parameter
name
Default
settings
Description
4.3.1 out
K-1
Gold alloy selection (K-1: gold / copper / silver;
K-2: gold / copper; K-3: gold / silver)
4.3.2
rtK
7 : 3
% ratio of copper and silver (in K-1 alloy).
Available settings – 4:6, 5:5, 6:4, 7:3
16.7.2. Table of gold alloy density
The table contains the density values of subsidiary components in an alloy
with regards to gold caratage in a tested sample: