Mode’s local settings, Table of platinum alloy density, Testing procedure – RADWAG WTB 600/AU Carat Balance User Manual
Page 42: Density, Table of platinum alloy

16.8.1. Mode’s local settings
Local settings of MODE P are grouped in submenu
No.
Parameter
name
Default
settings
Description
4.4 F–P
P-1
Testing platinum alloy – alloy selection
(P-1: platinum / nickel; P-2: platinum / palladium)
16.8.2. Table of platinum alloy density
The table contains the density values of subsidiary components in an alloy
with regards to platinum class in a tested sample:
Density of metal alloy [g/cm
3
]
Platinum
class [PT]
% content
Nickel Palladium
Range of sample
density [g/cm
3
]
1000
100 / 100
21.45
21.45
21.24 ~ 21.66
950
95.0 / 100
20.04
20.64
19.84 ~ 20.85
900
90.0 / 100
18.80
19.88
18.61 ~ 20.08
850
85.0 / 100
17.71
19.18
17.53 ~ 19.38
800
80.0 / 100
16.73
.18.53
16.56 ~ 18.72
750
75.0 / 100
15.86
17.92
15.70 ~ 18.10
16.8.3. Testing procedure
• Select working mode
>
in accordance with point 16.1 of this
user manual, after which the main window of selected working mode is
displayed:
• If measurement result is other than zero press
key (balance
zeroing).
- 42 -