• Load tested sample on the top weighing pan (weighing in the air,
flickering command IN AIR) and on stabilization of measurement
result press
key,
• Simultaneously the following command is displayed:
• Unload the tested sample from the top weighing pan,
• After the indication returns to ZERO place the tested sample in the
container with liquid (water) (weighing in liquid, flickering command
IN WATER) and on stabilization of measurement result press
key,
• Simultaneously, for approximately 2 seconds the software displays
liquid’s temperature value, and a window on process completion
appears (message END):
• The following test results are displayed on pressing
key.
Test results:
• Percent content of main metal in an alloy* [%],
• Density of tested sample [g/cm
3
],
• Volume of tested sample [cm
3
],
• Liquid
temperature
[
°C],
• Mass of weighed sample in the air [g],
• Mass of weighed sample in liquid (water) [g].